Focus shockwaves
principle
The principle is that the high-intensity focused shock wave energy generated by the shock wave generator enters specific deep tissues through the skin surface, effectively stimulates the affected tissue and accelerates the growth of new blood vessels and blood circulation, and induces a new inflammatory process, thereby awakening the cell's self-repair function. In addition, the effect of vaporization bubble burst generated by focused shock waves destroys the adhesion, scarring, contracture and blockage of the damaged tendon area, and even the calcified or fib rotic soft tissue, which is commonly known as loosening the injured part of the body, to stimulate the soft tissue to grow again, accelerate blood circulation and promote recovery. At the same time, high-energy shock waves can stimulate pain nerve receptors and block the conduction of pain signals, achieving an immediate analgesic effect.
Biologicaleffectsproducedbyshockwavesonthe humanbody
ApplicationfieldsofFocusshockwavesfocused shockwave
Indications
productadvantages
1. Intelligent operation and humanized design: The handle (hand piece) is
equipped with a 1.8-inch screen, making it easier to control and adjust on the handle, greatly improving the convenience and efficiency of operation.
2. Variable focusing and fine focus: With 7 kinds of replaceable silicone pad probes, fine and accurate penetration depth can be achieved. Suitable for pains of different depths and meets more treatment needs.
3. Deep penetration and wide coverage: By replacing silicone pad probes of different heights, this equipment can flexibly adjust the penetration depth of shock waves, making the penetration range of its treatment focus more accurate and meeting the treatment needs of different patients. Its strong penetration power enables deep tissues to be effectively treated as well, expanding the treatment range.
4. Non-invasive/non-surgical - no recovery time: As a non-invasive treatment method, Focus Shockwaves focused shock wave therapy does not require cutting the skin or injecting drugs, greatly reducing the patient's fear and discomfort and accelerating the rehabilitation process.
5. Safe and effective: It avoids the risks and complications of traditional surgeries and provides patients with a milder and more efficient recovery path.
6. Green therapy: With its safe, non-invasive and highly efficient characteristics, Focus Shockwaves focused shock wave therapy is known as a "green therapy", in line with modern medical pursuits of low damage and high recovery.
7. Comfortable treatment and painless on the body surface: Although piezoelectric shock waves have a relatively high peak pressure value at the focus, their wave source aperture angle is large and the energy density on the skin surface is low, reducing the pain on the skin surface and rarely causing tissue pain and discomfort.
8. Ultra-quiet and low noise: Whether it is a pneumatic ballistic type or other focused types, the noise generated during operation is relatively large and easily causes discomfort to patients. The piezoelectric focus shock wave adopts the anti-hemispherical surface integration technology and the piezoelectric element assembly to reduce frequency noise and realize a beam focusing device, providing a vibration-free and low-noise treatment environment.
9. The treatment time is short. Each treatment session lasts about 10 to 15 minutes, with quick effect and comfort.
InstrumentMaintenane&Maintenance
1. The instrument must use a plug with a grounding pin, and ensure that the power socket of the instrument is well grounded.
2. The power supply should be consistent with the specified value of the power supply marked on the machine, otherwise the machine may not work, or even burn out the mainboard parts of the machine.
3. Make sure that the power supply is stable and suitable. If the local power supply voltage is unstable, it is recommended that the user add a regulated power supply with matching power.
4. In order to ensure the therapeutic effect of the instrument and ensure the normal service life, please use the designated accessories provided or recommended by the original manufacturer.
5. When using the instrument, do not get close to the wall, and keep a space of 30cm around the instrument for heat dissipation.
6. The instrument is a high-precision electronic instrument, please do not place the instrument in a high temperature and humid environment.
7. The instrument adopts capacitive touch screen. When clicking, try to touch it with the pulp of your finger, and do not touch it with sharp objects.
8. Each accessory is a precision device, so it is necessary to pay attention to maintenance during and after use to avoid shortening its service life.
9. Do not use alcohol or corrosive solvents to clean the main unit and handle to avoid damage.
10. After each use of accessories, it is recommended to wipe and clean with clean water or normal saline, and then dry with a dry cloth.
11. The handle accessories must be handled with care during use, and should not be dropped or bumped by gravity to avoid damage to the handle accessories.The company will not provide warranty for damage to the handle accessories caused by artificial and violent bumps.
12. Patients will not be in contact with any metal or other alternate pathway to ground whilst the System is in use.
13. Before each use, please check whether the equipment and its accessories (especially the wires) are mechanically damaged or otherwise damaged.
14.When the instrument is not in use, please turn off the power, then pull out the power plug, and place various accessories of the instrument. If possible, please cover the instrument with a dust cover.
15. It is strictly forbidden to disassemble and modify the equipment without authorization.
16. If the equipment fails, it should be stopped immediately, please contact us.
warrantyInstructions
Host |
One year warranty |
Accessories |
Three months warranty |
1. Warranty period
(1)The warranty period of the product is calculated from the date of shipment from the factory.
(2)If any non-human damage or malfunction occurs within the warranty period, repair services will be provided free of charge.
2. The following situations are not covered by the warranty:
(1)WFaults caused by unauthorized maintenance, unauthorized d isassembly, modification, etc.;
(2)Damage caused by improper use, accidental falling or violent collision; ⑶Equipment failure caused by the use of non-original spare parts;
(3)Damage caused by natural disasters, force majeure, etc.
Listofproduction
Name |
Quantity |
Host |
1pc |
Handle |
1pc |
silicone pad |
7pcs |
Hanger |
2pcs |
Power Cable |
1pc |
Technicalspecifications
Product Name |
EM IMPACT POE |
Toughened glass capacitive touch screen |
15 Inches |
penetration depth |
4~66mm |
pulse frequency |
1~10Hz |
strength grade |
1~8Bar |
Operating Voltage |
AC100V~240V ,50Hz-60Hz |
Power |
10-500W |
Air box size |
53× 50×60cm |
Gross Weight |
23kg |
productInstructions
Introductiontosystem Interface
1. After booting up, enter the boot interface (Figure 1) and randomly click once to enter the interface shown in Figure 2.
Figure 1: Boot Interface
Treatmentparameterreferenceandcourse
1. Focus Shockwaves - Recommended usage range and parameters are as follows:
Surgical site |
silicone pad |
Number of impacts |
Frequency (Hz) |
Intensity (Bar) |
Shoulder (Calcium area of aponeurosis) |
No.3、No.4、No.5 |
1500~2000 Shots |
4~5Hz |
4~7 |
Scapulohumeral periarthritis |
No.3、No.4、No.5 |
1500~2000 Shots |
4Hz |
3~5 |
Elbow (Tennis elbow, Golfer's elbow) |
No.2 、No.3 |
1500~2000 Shots |
4~5Hz |
2~5 |
Wrist (carpal tunnel syndrome / tenosynovitis/TFCC syndrome) |
No.1 、No.2 |
1500~2000 Shots |
4Hz |
2~5 |
Thigh (iliotibial band syndrome/hamstring strain) |
No.5 、No.6 |
1500~2000 Shots |
4Hz |
2~5 |
Knee (Patellar chondromalacia / Patellar tendinitis) |
No.2 、No.3 |
1500~2000 Shots |
4Hz |
2~5 |
Front calf (Tibial anterior muscle syndrome /Tibial medial stress syndrome) |
No.1 、No.2 |
1500~2000 Shots |
4Hz |
2~5 |
Rear calf (Gastrocnemius muscle spasm) |
No.3 、No.4 |
1500~2000 Shots |
4Hz |
2~5 |
Foot (Achilles tendinitis) |
No.2 、No.3 |
1500~2000 Shots |
3Hz |
2~6 |
Foot (Calcaneal bone spur) |
No.3 、No.4 |
1500~2000 Shots |
4~5Hz |
4~7 |
Foot sole Plantar fasciitis |
No.2 、No.3 |
1500~2000 Shots |
4Hz |
3~7 |